SPECIALTY GAS MANIFOLDS - HBAC2 AUTOMATIC MANIFOLD
specs HBAC2 Technical Data Sheet  installation & operation
High Purity Brass Manifolds:
 
• Stainless steel diaphragm regulators.
• Helium leak rate integrity 2 x 10-6 scc/sec.
• Maximum inlet pressure 3000 psig. (2000 psi for CO2 & N2O systems)
• Maximum flow rate:
HBAC2 250 scfh
•Delivery pressure range:
HBAC2 30–100 psig
Acetylene 0–15 psig
HBAC2HP 50–200 psig
• Manifold outlet: 1/4” OD tube.
• 1/2” Brass, silver brazed headers
• 24” flexible stainless steel lined pigtails with check valves.


Western Innovator HBAC2 manifolds are designed and manufactured for high-purity gas delivery applications requiring uninterrupted gas flow and greater cylinder capacities. User friendly and easy to operate, a simple rotation of the control knob resets the unit.

Factory-set functional components are protected inside a tamper-resistant case. A self-contained alarm system clearly indicates the system status. A green light means the service bank is functioning and the reserve bank is ready for service. A red light alerts the user that the unit has changed over and one or both banks are depleted. Dry contacts in the unit’s power supply box allow connection to remote alarms systems. (Up to 3 amps 30 VDC or 2 amps 250VAC).

high purity brass manifold


 



Design Lengths HBAC2
TOTAL NO. OF CYLINDERS
2
4
6
8
10
12
16
20
HBAC2 Std (10” Centers)
Overall Manifold Length
2’-3”
(.7m)
3’-11”
(1.7m)
5’-7”
(2.21m)
7’-3”
(2.72m)
8’-11”
(2.95m)
10’-7”
(3.23m)
13’-11”
(4.24m)
17’-3”
(5.26m)
HBAC2 Staggered Design (5” Centers)
Overall Manifold Length
2’-3”
(.7m)
3’-1”
(.94m)
3’-11”
(1.2m)
4’-9”
(1.45m)
5’-7 ”
(1.7m)
6’- 5 ”
(1.96m)
8’-1”
(2.46m)
9’-9”
(2.97m)
HBAC2 Vertical Crossover (10” Centers)
Overall Manifold Length
N/A
2’-3”
(.7m)
N/A
3’-11”
(1.2m)
N/A
5’-7”
(1.7m)
7’-3”
(2.21m)
8’-11”
(2.72m)
HBAC2 Standard (13” Centers) (Acetylene)
Overall Manifold Length
2’-3”
(.7m)
4’-5”
(1.35m)
6’-7”
(2m)
8’-9”
(2.7m)
10’-11”
(3.3m)
13’-1”
(4m)
17’-5”
(5.31m)
21’-9”
(6.63m)
HBAC2 Staggered Design (6.5” Centers)
(Acetylene) Overall Manifold Length
2’-3”
(.7m)
3’-4”
(1m)
4’-5”
(1.35m)
5’-6”
(1.67m)
6’-7”
(2m)
7’-8”
(2.34m)
9’-10”
(3.00m)
12’-9”
(3.66m)
HOW TO ORDER
Specialty Manifold Ordering Instructions
Specify: Control Type (V) - Service (W) - Number
Example : HBAC2–5–6 = Model HBAC2, Helium service
CONTROL TYPE (V) GAS SERVICE (W) # OF
CYL’S (X)
HEADER
CONFIGURATION (Y)
MOUNTING (Z)
HBAC2 (30 - 100 PSIG)

HBAC2HL (30 - 100 PSIG)
(includes 500 SCFH heater)

HBAC2HP
(50 - 200 PSIG)
(1) Acetylene (POL)* CGA-510
(1A) Acetylene (Commercial)* CGA-300
(2) Zero Air CGA-346
(2A) Zero Air CGA-590
(3) Argon CGA-580
(4) Carbon Dioxide CGA-320
(5) Helium CGA-580
(6) Hydrogen CGA-350
(6A) Argon/Methane CGA-350
(7) Nitrogen CGA-580
(8) Nitrous Oxide CGA-326
(9) Oxygen CGA-540
  BLANK - Standard 10 inches on center
13 inches on center for Acetylene & LPG
V - Vertical crossover
Standard 10 inches on center
13 inches on center for Acetylene & LPG
S - Staggered 5 inches on center
6.5 inches on center for Acetylene & LPG
C - Crossover (Floor Mount Only)
Standard 10 inches on Center
13 inches on center for Acetylene & LPG
U-Shaped - Drawing Required
L-Shaped - Drawing Required
BLANK = Wall mount
F = Floor mounted
*Not available for HBAD

For more information, call Western Customer Service
at 1-800-783-7890 for technical data sheets

 

 


 

home | about us | distributor info | wholesalers | service providers
industrial manifolds | med manifolds | spec manifolds | manifold accessories
news | faqs | references


 
 
powered by Vertical Lift Web Designs
copyright © 2007 Western Innovator, Inc.